Furnace Structure
| - The refractory muffle is built-in with a water-cooling wall with dimension 470 LX 410W X 550H (mm), which can stand at positive pressure up to 0.06Mpa and vacuum up to - 0.09Mpa.
- Built-in pressure gauge, vacuum valves, and one gas float meters
- MoSi2 heating elements for working in both inert gas and oxygen environment up to 1600oC
- Two digital temperature controllers for controlling and monitoring furnace running at safety
|
Inside Chamber
| - Heated chamber: 400x 400x 400 mm ( 16x16x16 inch), 65 Liter ( Max. )
- Energy-saving 1800-grade fibrous alumina insulation
- Sealed by stainless steel with water cooling
- Two sides heating ( Left and right ) to ensure the best temperature uniformity.
- SiC plate is installed on the bottom, which has 100kg, Max. loading
|
Power | - AC 208 - 240V, three phases (50/60 Hz). 400 or 480VAC is also available upon request.
- 36 KW
- 80 Amp (100 Amp air breaker is required)
- 9' Power cable is included without a plug. Suggest you install the power plug according to your lab power panel by a local electrician.
|
Working Temperature & Heating Rate | - Continuous: 800°C - 1600°C
- Max. 1650°C (< 2 hr)
- Heating Rate: 10°C / min max.
- Temp. Accuracy: +/- 1°C
|
Heating Elements
| - 12 sets of MoSi2 coils heating elements are built-in both sides of the chamber and replaceable
- The heating element is consumable parts that should be ordered separately if an extra set of heating elements is required.
- MoSi2 heating elements can be heated both in inert gas and air
|
Temperature Controllers
| - Two temperature controllers and thermocouples (T/C) are built-in the furnace
- The primary controller and thermocouple dominate the user-defined heating program and the main alarm system with +/- 1oC accuracy and 30 segments of programable
- The monitoring controller only monitors the real temperature by the other thermocouple and activates the secondary alarm system when the failure of primary controller to ensure safety.
|
Gas Flow Control and Vacuum
| - Gas Flow Control and Vacuum Valve one large flowmeter (12L/min) is installed in the front panel o
- KF40 vacuum port with the valve is built-in to connect a vacuum pump
- The Heavy-duty vacuum pump is included. Please click the Pic. left to see specs
|
Water Cooling
|
- A 16L/min recirculating water chiller is included for the cooling air-tight chamber and the front door.
|
Overall Dimensions
| |
Shipping Weight | |
Warranty
| - One-year limited warranty with lifetime support. (Consumables including heating elements and O-rings are not covered by the warranty, please order the replacement at related products below.)
|
Compliance | - CE Certified
- NRTL or CSA certification is available upon request at the extra cost.
|