Mixing
| - Planetary Centrifugal Vacuum Mixing
- The centrifugal force of rotation and revolution induces convection, homogenizing the materials
- With the vacuum pressure, submicron air bubbles can be efficiently eliminated from the materials
|
Containers and Adapters | Please choose the desired container and the adapter from the option bar.
- Option 1: includes one 300ml container with gas vent, made of high-density Polyethylene or PTFE
- Option 2: includes one 100ml adapter made of high-density Polyethylene or PTFE
- Option 3: includes one 100ml container with gas vent, made of high-density Polyethylene or PTFE
- Option 4: includes one 100ml adapter and one 30ml adapter made of high-density Polyethylene or PTFE
- Option 5: includes one 30ml container with gas vent, made of high-density Polyethylene or PTFE
- Option 6: includes one 10ml container adapter, made of high-density Polyethylene or PTFE
Note: Both 30ml and 100ml adapters are needed for using the 30ml containers. Please click the pictures below for more detail. 100ml Container 30ml Container 3x10ml Container |
Dimensions
| - 300ml containers: 76mm (O.D.) x 88mm (Height)
- 100ml containers: 60mm (O.D.) x 57mm (Height)
- 30ml containers: 40mm (O.D.) x 35mm (Height)
- 100ml adapters: 83mm (O.D.) x 61.75mm (I.D.) x 35mm (Height)
- 30ml adapters: 68mm (O.D.) x 40mm (I.D.) x 28.5 (Height)
300ml Container 100ml Container 30ml Container 100ml Adapter 30ml Adapter |
Maximum Mixing Volume and Weight |
Container | Max Sample Volume | Max Sample Weight | Max Gross Weight (Container Included) | 300 mL | 150 mL | 140 g | 200 g | 100 mL | 50 mL | 45 g | 80 g (adapter not included) | 30 mL | 15 mL | 12 g | 20 g (adpater not included) |
- Attention: Sample material overload might damage the motor and/or cause spilling. Follow both the volume and weight limit when loading the sample
|
Application Notes | - Always use the mixer on a level & stable surface for best performance and maximum safety
- Do not use the mixer in a hazardous atmosphere, or with flammable, explosive, corrosive, poisonous material! Check the chemical resistance of the container with the different materials before loading the sample!!!
- Do not overload the mixer! Stop the mixing operation immediately if abnormal sound/vibration are observed
- Use a digital balance to make sure the sample loading is balanced before mixing
- Double-check if the containers are mounted correctly and securely before mixing
- Do not open the cover lid until the rotation has completely stopped! Do not retrieve the containers and/or put hand close to the mixing fixture while the rotation has not completely stopped!
- Only use the included containers for the mixing operation!
|