Voltage | 220V, 50Hz (110V is available with 2000W transformer)
|
Power | 1500W
|
Hydraulic Cylinder Pressure | 0 - 40T for 30 mm ID vessel 0 - 110T for 50mm ID vessel |
Bottom Cylinder (Piston) Traveling Distance | 50 mm Maximum |
CIP Vessel Size | • Inside: 30 mm (I.D.) x 110 mm (H) Overall: Φ100 × 315 H (mm) Weight: 20Kg- • Inside: 51mm (I.D.) x 150mm (H) Overall: Φ180 × 350 H (mm) Weight: 50 Kg
-
|
Vessel Pressure
| • 0 - 500 MPa- • For calculating the actual pressure (MPa) applied on your working piece, please follow the
equations below: - • P = T/πr^2 x g x 10
- • P: Pressure (Mpa) applied to your working piece
- • T: Force (metric ton) Applied on the Pressure Vessel Cylinder
- • g: Standard gravity. g ≈ 9.8 m/s2
- • π: ~3.14
- • r: Radius (cm) of the Pressure Vessel ( 1.5 cm)
- • *1000 Kg = 1 Metric Tone; 1 Mpa = 10 Kg/cm2
|
Dimension of Machine
| 400mm (L) x 450mm (W) x 1100mm (H) for 110T Press |
Net Weight | 580 Kg
|
Compliance | CE Certified |
Warranty | One year |
Operation Video & Instructions |
Operation Video O-Ring Installation
|
Application Notes |
|
Warning | • The Max. pressure the mold can handle is 85T, please do NOT exceed it- • Never let the press holding the target pressure for more than 8 minutes
- • It's normal that the target pressure decreases a little bit during the pressure holding time
- • Pressure must be completely released first, then make sure the hand wheel and the pressure
release valve has been tightened before pressing - • Do not raise the bottom cylinder (piston) more than it's maximum traveling distance (50mm)
|