EQ-GSL-2000X-25 is an ultra-high temperature furnace with 1" I.D x 6" L heating chamber up to 2000 °C under inert gas for materials research. This compac furnace uses a special tungsten heating element design to reach high temperatures at very low power (< 3000 W) . Precision digital controller provides stable temperature control via C Type thermocouple and SCR power supply.
SPECIFICATIONS
Work Volume
1 zone
Cylindrical heated volume
•25 mm diameter, •150 mm length
Operating Temperature
•2000°C Maximum (short time < 30 minutes) under high purity (>=99.999%)1 atm Argon gas. •1900°C Max. for continuous under high purity (>=99.999%)1 atm Argon gas.
Temperature Stability
+/- 1 °C
Maximum Ramp Rates
•Heating Rate
•10 °C/minute typical; •5 °C/min at high temperatures
•Cooling Rate
•10 °C/minute typical; •Slow/strongly temperature dependent at low temperatures
Operating Atmosphere
•1 atm inert gas. or reducing gas at temperatures below approximately 2000 °C. •Not suitable for high vacuum
•Type C thermocouple inserted from the furnace top •Two C T/C included
Temperature Control
•30 segment Digital Program Controller provides a 4-20 mA signals to a SCR power supply. •Built in protection from Over-temperature and thermal couple broken •+/- 1 °C temperature accracy •PC control software anid interface is available at extra cost
Vacuum Fittings
•Main chamber sealed with silicone rubber O-ring.
Outer seals standard KF seals. •KFD25 vacuum port and gas inlet & outlet valve are built in •Vacuum guage installed
Water Cooling
•Water cooled two end caps •30 L/min water flow required
Sample holder
•Sample holder is driven by electric motor , and automatically move up and dow to make sample loading easy. •one 18 mm O.D x 16 mm I.D tungsten disc is installed on bottom thermal block for holding a crucible or sample directly
Power Supply
SCR power supply, 20 A, VAC 208 - 240, 50/60Hz,
single phase power supply
4000 W Max. 3000 W operation
Utility Requirements
•Furnace Electrical Supply : 220/240 volt AC, single phase, 20 Amp – 50/60 Hertz, two-wire plus ground. •Mechanical or molecular pump is required before purging inert gas. You may order at related product below •Nominal 4 Liters per minute (lpm) of cooling water. You may tap water or order our Recirculating Water Chiller 58 L/min Flow, ( order at related product below )
Overall Dimensions
800 mm L × 750 mm W × 1600 mm H
Warranty
One year limited with lifetime support (Consumable parts such as processing tubes, o-rings and heating elements are not covered by the warranty, please order the replacement at related products below.)
Compliance
•CE Certified •UL / MET / CSA Certification is available upon request, additional charges will apply •Temperature controller is MET and CE Certified •Transformer is UL Certified
One year limited with lifetime support (Consumable parts such as processing tubes, o-rings and heating elements are not covered by the warranty, please order the replacement at related products below.)