PTL-SC-6S-LD Coating System has been designed to automate the entire process to avoid operator fatigue and errors associated with it.
In manual SILAR process, the operator has to perform hundreds of repetitive dipping into the solution and water.
It is very difficult to control dip duration and number of dips in a manual process which can last hours.
In the automated unit, the operator need just to clamp the substrate into the holder and program the controller with required dip cycles and duration.
PTL-SC-6 has a motorized substrate holder .
The dipping speed, dip duration, retrieval speed and dry duration can be set for each beaker.
Each hot plate can be set at different temperatures.
The temperature can be set up to 350°C from ambient.
In this technique thin film is deposited on glass substrate following a chemical technique called successive ion layer adsorption and reaction (SILAR).
The technique involves multiple dipping of the substrate in a given solution and deionized water, temperature of both can vary from case to case.
SPECIFICATIONS
Actuator
Stepper motor
Drive Mechanism
Lead screw
Dip Duration
0 - 99 seconds / minutes / hours
Speed Control
Drawing speed adjustable from digital pannel from 2 micron/sec to 9000 micron/sec ( Max.)
Number of Dips
1 - 999
Hot Plate Temperature
Ambient to 350 °C
Stirrer Speed
1 - 200 rpm
Power Input
•110 V or 208 - 240 VAC selectable, 50 / 60 Hz •600 W max ( with heater )
Stroke length Max.
75 mm
Control Unit
•Speed control via LED display with 5 programs memory •Temperature control by 6 PID digital controller with +/- 1 °C accuracy upto 350 °C •RS232 port for connect PC •Opertion software included
Product Dimensions
Certificate
CE Certified
Warranty
One year limited with lifetime support
Operation Instructions
Application note
•In order to get better film coating, the substrate shall be treated by plasma cleaner (Please click on the first picture below for more information) •Use ultrasonic processor to disperse nano powder or homogenize solution to get better coating (Please click on the second picture below for more information) •The compact dip coater can be put in a MTI glovebox to operate under control atmosphere •Click here to visit our Knowledge Center and learn about dip coating