Power
| - 208 - 240 VAC / 50/60Hz, 1 ph, 8 kW (50 A power line is required)
- 10-ft power cord
- A power plug is not included at an extra cost.
|
Ultimate Vacuum | |
Working Temperature | |
Heating Rate | |
Heating Area | - Dual heating zone
- ID 250 × H 350 mm (175+175mm)
|
Effective Heating Area | - Ø200 × H350 mm (maximum sample size)
|
Heating Chamber | - 310 stainless steel Ø216×Ø206×600H
- DN200CF flange
|
Thermocouple | |
Heating Elements | |
Thermal Block | |
Temperature Controller
| - Two EPC3004 Eurotherm PID temperature controllers
- 28 temperature segments
- ±0.1℃accuracy/resolution
- RS485 communication port
|
Chilled Water
| - Two Ø12 mm push-to-connect water inlet and outlet.
- Recommended cooling capacity: 1500 – 2500 btu/h
- Recommended water flow: 5-15L/min
|
Overall Dimensions
| 650mm(L)× 650mm(W)× 1250mm(H) |
Net Weight | 100 kg |
Shipping Weight | 250 lbs |
Warranty
| - One-year limited warranty
- Consumables are not covered, including the SS heating chamber, heating elements, thermocouples, and thermal blocks.
- Any damage caused by misuse or negligence or use of corrosive gas are not covered.
|
Compliance
| - CE Certified
- NRTL or CSA certification is available upon request at extra cost.
|
Warning | - The heating chamber is compatible with vacuum and ambient pressure. The chamber pressure should NOT exceed the ambient pressure. Never use the furnace at a positive pressure.
|
Application Notes | - Turbo pump Water Chiller Quartz Block Copper Gasket
-
|
Reference Video | |