Furnace Structure | - Double shell steel case with a cooling fan to keep case surface cool
- High-purity alumina fiber insulation for max. energy saving
- ZrO2 (YSZ) chamber liner for longer service life
|
Dimensions | - Chamber: 200x 200x 300 mm ( 8 x 8 x 12 inch), 12 Liter
- Overall: 600 x 780 x 620 mm (24 x 31 x 24.5 inch)
|
Power | AC 208V-240V single phase, 50/60 Hz (50A air breaker required 5KW |
Standard Working Temperature | 1750°C (continuous) |
Maximum Working Temperature | 1800°C (< 1 hours) |
Heating Rate | - RT ~10°C/min ( RT - 1600°C )
- 2 ~5 °C/min (1600 ~ 1800°C )
- Note: lower heating rate will result in the longer service life of heating elements
|
Temperature Accuracy | +/- 1° C |
Temperature Uniformity | - +/- 5°C over 120 mm (5") @ 1700°C
- +/- 2°C over 80 mm (3") @ 1700°C
|
Heating Elements | 8 pcs, Kanthal Super-1900 grade MoSi2 Heating Element - 420mm L, |
Thermocouple | B-Type (Pt-Rh to Pt-Rh) with YSZ tube |
Temperature Controller
| - FA-YD518P-AG temperature controller is included.
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- Default DB9 PC communication port
|
Temperature controller (optional)
| - MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for detailed information about the MTS02-Y software kit.
- You may upgrade the temperature controller to the Eurotherm 3504 Temperature Controller which is fully compatible with LabVIEW. This package includes a Eurotherm 3504 programmer, a communication cable (RS485 - USB), and a software CD. Click the pic below to view the full spec of the Eurotherm 3504 temperature controller.
|
Laptop, software(Optional)
| - Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)
- Labview-Based Temperature Control System (EQ-MTS01) enables users to edit temperature profiles, manage heat-treatment recipes, and record and plot data for MTI furnaces.
- A wireless remote control provides up to 300 meters of operating range
|
Power Cable & Plug | - 10 feet long 10-3 AWG heavy duty UL approved power cable is included
- Extra length cable can be ordered by clicking the picture below
- The plug is not included. You can order from your local hardware store according to the female plug in your lab, or order from the link below (30A plug)
(click here to order plugs from us) |
Net Weight | 160Kg |
Shipping Weight | 420lbs |
Shipping Dimensions | 34" x 32" x 53" |
Warranty | - One-year limited warranty with lifetime support. (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.)
- ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of MTI One Year Limited Warranty.
|
Compliance | - CE Certified
- NRTLor CSA certification is available at an extra cost.
|
Application Notes | - For preserving the longevity of the furnace, we recommend the heating rate not exceed 10°C /min.
- The cooling rate shall also not exceed 10 °C /min.
- This furnace lacks vacuum sealing, and toxic or explosive gases are not recommended for use with this furnace without necessary safety controls and supervision.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
- Please click here to learn how to DIY a gas purging system.
- The refractory block must be inserted before closing the door.
|
Operation Instructions |
|